Verification and Calibration - B
57
Table B- 3. Verification Test Record for Agilent 6631B
Model Agilent 6631B Report No.____________ Date_____________
Test Description Minimum
Specification
Recorded Results Maximum
Specification
Voltage Programming and Measurement
Low Voltage V
out
5 mV
_______V +5 mV
Front Panel measurement
V
out
2 mV
_____mV V
out
+2 mV
High Voltage V
out
8.991 V _______V 8.009 V
Front Panel measurement
V
out
4.4 mV
_____mV V
out
+4.4 mV
Current Programming and Measurement (High Range)
Low Current I
out
4 mA
_______A 4 mA
High Current I
out
9.991 A _______A 10.009 A
Front Panel measurement
I
out
21 mA
_____mA I
out
+21 mA
Current Measurement (Low Range)
Low Current measurement
I
out
2.5
µ
A _____
µ
AI
out
+2.5
µ
A
High Current measurement
I
out
22.5
µ
A _____
µ
AI
out
+22.5
µ
A
Current Sink Measurement
Low Current Sink measurement
I
sin
k
22.5
µ
A _____
µ
AI
sin
k
+22.5
µ
A
High Current Sink measurement
I
sin
k
21.6 mA
_____mA I
sin
k
+21.6 mA
Table B- 4. Verification Test Record for Agilent 66332A or Agilent 6632B
Model Agilent______________________ Report No.____________ Date_____________
Test Description Minimum
Specification
Recorded Results Maximum
Specification
Voltage Programming and Measurement
Low Voltage V
out
10 mV
_______V +10 mV
Front Panel measurement
V
out
3 mV
_____mV V
out
+3 mV
High Voltage V
out
19.980 V _______V 20.020 V
Front Panel measurement
V
out
9 mV
_____mV V
out
+9 mV
Current Programming and Measurement (High Range)
Low Current I
out
2 mA
_______A 2 mA
High Current I
out
4.9955 A _______A 5.0045 A
Front Panel measurement
I
out
10.5 mA
_____mA I
out
+10.5 mA
Current Measurement (Low Range)
Low Current measurement
I
out
2.5 µA _____µAI
out
+2.5 µA
High Current measurement
I
out
22.5 µA _____µAI
out
+22.5 µA
Current Sink Measurement
Low Current Sink measurement
I
sin
k
22.5 µA _____µAI
sin
k
+22.5 µA
High Current Sink measurement
I
sin
k
11.1 mA
_____mA I
sin
k
+11.1 mA

Table of Contents

Other manuals for Agilent Technologies 6632B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6632B and is the answer not in the manual?

Agilent Technologies 6632B Specifications

General IconGeneral
BrandAgilent Technologies
Model6632B
CategoryPower Supply
LanguageEnglish

Related product manuals