Verification and Calibration - B
57
Table B- 3. Verification Test Record for Agilent 6631B
Model Agilent 6631B Report No.____________ Date_____________
Test Description Minimum
Specification
Recorded Results Maximum
Specification
Voltage Programming and Measurement
Low Voltage V
out
−
5 mV
_______V +5 mV
Front Panel measurement
V
out
−
2 mV
_____mV V
out
+2 mV
High Voltage V
out
8.991 V _______V 8.009 V
Front Panel measurement
V
out
−
4.4 mV
_____mV V
out
+4.4 mV
Current Programming and Measurement (High Range)
Low Current I
out
−
4 mA
_______A 4 mA
High Current I
out
9.991 A _______A 10.009 A
Front Panel measurement
I
out
−
21 mA
_____mA I
out
+21 mA
Current Measurement (Low Range)
Low Current measurement
I
out
−
2.5
µ
A _____
µ
AI
out
+2.5
µ
A
High Current measurement
I
out
−
22.5
µ
A _____
µ
AI
out
+22.5
µ
A
Current Sink Measurement
Low Current Sink measurement
I
sin
−
22.5
µ
A _____
µ
AI
sin
+22.5
µ
A
High Current Sink measurement
I
sin
−
21.6 mA
_____mA I
sin
+21.6 mA
Table B- 4. Verification Test Record for Agilent 66332A or Agilent 6632B
Model Agilent______________________ Report No.____________ Date_____________
Test Description Minimum
Specification
Recorded Results Maximum
Specification
Voltage Programming and Measurement
Low Voltage V
out
−
10 mV
_______V +10 mV
Front Panel measurement
V
out
−
3 mV
_____mV V
out
+3 mV
High Voltage V
out
19.980 V _______V 20.020 V
Front Panel measurement
V
out
−
9 mV
_____mV V
out
+9 mV
Current Programming and Measurement (High Range)
Low Current I
out
−
2 mA
_______A 2 mA
High Current I
out
4.9955 A _______A 5.0045 A
Front Panel measurement
I
out
−10.5 mA
_____mA I
out
+10.5 mA
Current Measurement (Low Range)
Low Current measurement
I
out
−2.5 µA _____µAI
out
+2.5 µA
High Current measurement
I
out
−22.5 µA _____µAI
out
+22.5 µA
Current Sink Measurement
Low Current Sink measurement
I
sin
−22.5 µA _____µAI
sin
+22.5 µA
High Current Sink measurement
I
sin
−11.1 mA
_____mA I
sin
+11.1 mA