B - Verification and Calibration
58
Table B- 5. Verification Test Record for Agilent 6633B
Model Agilent 6633B Report No.____________ Date_____________
Test Description Minimum
Specification
Recorded Results Maximum
Specification
Voltage Programming and Measurement
Low Voltage V
out
20 mV
_______V +20 mV
Front Panel measurement
V
out
6 mV
_____mV V
out
+6 mV
High Voltage V
out
49.955 V _______V 50.045 V
Front Panel measurement
V
out
21 mV
_____mV V
out
+21 mV
Current Programming and Measurement (High Range)
Low Current I
out
1 mA
_______A 1 mA
High Current I
out
1.998 A _______A 2.002 A
Front Panel measurement
I
out
4.25 mA
_____mA I
out
+4.25 mA
Current Measurement (Low Range)
Low Current measurement
I
out
2.5
µ
A _____
µ
AI
out
+2.5
µ
A
High Current measurement
I
out
22.5
µ
A _____
µ
AI
out
+22.5
µ
A
Current Sink Measurement
Low Current Sink measurement
I
sin
k
22.5
µ
A _____
µ
AI
sin
k
+22.5
µ
A
High Current Sink measurement
I
sin
k
4.85 mA
_____mA I
sin
k
+4.85 mA
Table B- 6. Verification Test Record for Agilent 6634B
Model Agilent 6634B Report No.____________ Date_____________
Test Description Minimum
Specification
Recorded Results Maximum
Specification
Voltage Programming and Measurement
Low Voltage V
out
50 mV
_______V +50 mV
Front Panel measurement
V
out
12 mV
_____mV V
out
+12 mV
High Voltage V
out
99.90 V _______V 100.10 V
Front Panel measurement
V
out
42 mV
_____mV V
out
+42 mV
Current Programming and Measurement (High Range)
Low Current I
out
0.5 mA
_______A 0.5 mA
High Current I
out
0.999 A _______A 1.001 A
Front Panel measurement
I
out
2.25 mA
_____mA I
out
+2.25 mA
Current Measurement (Low Range)
Low Current measurement
I
out
2.5 µA _____µAI
out
+2.5 µA
High Current measurement
I
out
22.5 µA _____µAI
out
+22.5 µA
Current Sink Measurement
Low Current Sink measurement
I
sin
k
22.5 µA _____µAI
sin
k
+22.5 µA
High Current Sink measurement
I
sin
k
2.85 mA
_____mA I
sin
k
+2.85 mA

Table of Contents

Other manuals for Agilent Technologies 6632B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6632B and is the answer not in the manual?

Agilent Technologies 6632B Specifications

General IconGeneral
BrandAgilent Technologies
Model6632B
CategoryPower Supply
LanguageEnglish

Related product manuals