The number visible in the Current Position box is properly stored in the
appropriate location.
Repeat these steps to properly set the manifold sampling depth for any
subsequent apparatus (i.e., 1, 2, 5, or 6) / volume combination.
5.3. AutoTemp
The AutoTemp option of the 708-DS dissolution apparatus includes temperature
probes at each active vessel location. This enables the analyst to have
documented temperature values at the beginning and end of the dissolution
method as well as any pre-programmed sample time points. Basic use of this
option is described in Section 3.2.2.
5.3.1. Enabling Temperature Measurements
To enable initial and final vessel temperature measurements, place a
check in the appropriate boxes from the Method Properties screen shown
in Figure 5-4.
Figure 5-4. Method Properties Screen 3
Details of how to setup the method are discussed in Section 6 of this
manual. Section 3.2.2 offers details regarding the recommended manifold
lead time and duration for vessel temperature measurements at individual
sample time points.
5.3.2. Setting Tolerances
Programmable tolerances for various method parameters are defined
within the firmware of the apparatus. These are set up from screen 2 of
the Instrument Settings. Detailed explanation on how to navigate to this
screen and enter the values is discussed in Section 2.2.4.