7-6
Specifications and Regulatory Information
Specifications
Table 7-2. Resolution Bandwidth (RBW) Specifications
Resolution Bandwidth (RBW)
Agilent 86140B,
86142B, 86143B,
86145B
Agilent 86144B,
86146B
a
a. The 86144B and 86146B specifications are for the 50 µm internal path only.
Agilent 86141B,
86140B w/Opt 025,
86143B w/Opt 025
FWHM (3 dB Bandwidth)
b
,
c
b. With applied input fiber 9/125 µm
c. Resolution of 10 nm is available for first order grating response only
0.06, 0.1, 0.2, 0.5, 1, 2, 5,
10 nm
0.06, 0.07, 0.1, 0.14, 0.2,
0.33, 0.5, 1, 2, 5, 10 nm
0.07, 0.1, 0.2, 0.5, 1, 2,
5, 10 nm
Noise Marker Bandwidth Accuracy
using noise markers 1525-1610 nm
a,
d
0.5 nm
0.2 nm
0.1 nm
0.06 nm
d. Temperature range 20 to 30°C
±2%
±3%
±7%
±12%
2%
±3%
±7%
±12%
±3%
±5%
±10%
-----