7-7
Specifications and Regulatory Information
Specifications
Table 7-3. Amplitude Specifications
Amplitude
Agilent 8614xB
a
Sensitivity
b
600-750 nm
c
,
d
750-900 nm
c.d
900-1250 nm
c
1250-1610 nm
c
1610-1700 nm
e
60 dBm
75 dBm
75 dBm
90 dBm
80 dBm
Maximum Measurement
Power
f
,
g
1525-1700 nm
600-1000 nm
1000-1525 nm
+15 dBm per channel,
+30 dBm total
+15 dBm per channel,
+30 dBm total
+12 dBm per channel,
+30 dBm total
Maximum Safe Power
Total safe power
Total power within any 10 nm
portion of the spectrum
+30 dBm
+23 dBm
Absolute Accuracy
h
at 20 dBm, 1310 nm/1550 nm
±0.5 dB
Scale Fidelity
e,
i
Autorange off
± 0.05 dB
Autorange on
± 0.07 dB
Display Scale (log scale) 0.01-20 dB/DIV.
120 to +90 dBm
Amplitude Stability
(1310 nm, 1550 nm)
1 minute
j
15 minutes
g
±0.01 dB
±0.02 dB