7-7
Specifications and Regulatory Information
Specifications
Table 7-3. Amplitude Specifications
Amplitude
Agilent 8614xB
a
Sensitivity
b
600-750 nm
c
,
d
750-900 nm
c.d
900-1250 nm
c
1250-1610 nm
c
1610-1700 nm
e
60 dBm
75 dBm
75 dBm
90 dBm
80 dBm
Maximum Measurement
Power
f
,
g
1525-1700 nm
600-1000 nm
1000-1525 nm
+15 dBm per channel,
+30 dBm total
+15 dBm per channel,
+30 dBm total
+12 dBm per channel,
+30 dBm total
Maximum Safe Power
Total safe power
Total power within any 10 nm
portion of the spectrum
+30 dBm
+23 dBm
Absolute Accuracy
h
at 20 dBm, 1310 nm/1550 nm
±0.5 dB
Scale Fidelity
e,
i
Autorange off
± 0.05 dB
Autorange on
± 0.07 dB
Display Scale (log scale) 0.01-20 dB/DIV.
120 to +90 dBm
Amplitude Stability
(1310 nm, 1550 nm)
1 minute
j
15 minutes
g
±0.01 dB
±0.02 dB

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 86140B Series and is the answer not in the manual?

Agilent Technologies 86140B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model86140B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals