30 G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
7 Select Full Quant from Method Tasks section 4 to display the FullQuant
Method Table pane:
8 Set the following calibration curve parameters:
The calibration curve method (External Calibration / Standard Addition)
Type of VIS correction for ISTD elements
Curve fit type (Linear / Quadratic / Excluded)
Handling of the origin (Ignore / Force / Blank offset)
Calibration curve weight
ISTD elements used for correction
Concentration of the calibration curve level (Min Conc). For most
applications, enter 0.
Concentration unit
Concentration of the blank