EasyManuals Logo
Home>Agilent Technologies>Power Supply>E3633A

Agilent Technologies E3633A User Manual

Agilent Technologies E3633A
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #101 background imageLoading...
Page #101 background image
Chapter 4 Remote Interface Reference
Calibration Commands
88
CALibration:VOLTage:LEVel {MINimum | MIDdle|MAXimum}
This command can only be used after calibration is unsecured and the output state is
ON. It sets the power supply to a calibration point that is entered with CAL:VOLT
command. During calibration, three points must be entered and the low-end point
(MIN) must be selected and entered first.
CALibration:VOLTage:PROTection
This command calibrates the overvoltage protection circuit of the power supply. It
takes about 10 seconds to execute the command. The calibration must be unsecured
and the output be opened before calibrating the overvoltage protection circuit. The
power supply automatically performs the calibration and stores the new overvoltage
constant in nonvolatile memory. Notice that voltage calibration precedes before
sending this command.

Table of Contents

Other manuals for Agilent Technologies E3633A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3633A and is the answer not in the manual?

Agilent Technologies E3633A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3633A
CategoryPower Supply
LanguageEnglish

Related product manuals