Chapter 4 Remote Interface Reference
An Introduction to the SCPI Language
103
4
Command Separators
A colon ( : ) is used to separate a command keyword from a lower-level keyword as
shown below:
‘‘SOURce:CURRent:TRIGgered’’
A semicolon ( ; ) is used to separate two commands within the same subsystem, and
can also minimize typing. For example, sending the following command string:
‘‘SOUR:VOLT MIN;CURR MAX’’
... is the same as sending the following two commands:
‘‘SOUR:VOLT MIN’’
‘‘SOUR:CURR MAX’’
Use a colon and a semicolon to link commands from different subsystems. For
example, in the following command string, an error is generated if you do not use
the colon and semicolon:
‘‘DISP:TEXT:CLE;:SOUR:CURR MIN’’
Using the MIN and MAX Parameters
You can substitute MINimum or MAXimum in place of a parameter for many
commands. For example, consider the following command:
CURRent {<current>|MIN|MAX}
Instead of selecting a specific current, you can substitute MINimum to set the current
to its minimum value or MAXimum to set the current to its maximum value.

Table of Contents

Other manuals for Agilent Technologies E3633A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3633A and is the answer not in the manual?

Agilent Technologies E3633A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3633A
CategoryPower Supply
LanguageEnglish

Related product manuals