Chapter 4 Remote Interface Reference
Triggering Commands
80
To ensure synchronization when the bus source is selected, send the *WAI (wait)
command. When the *WAI command is executed, the power supply waits for all
pending operations to complete before executing any additional commands. For
example, the following command string guarantees that the first trigger is accepted
and is executed before the second trigger is recognized.
TRIG:SOUR BUS;*TRG;*WAI;*TRG;*WAI
You can use the *OPC? (operation complete query) command or the *OPC
(operation complete) command to signal when the operation is complete. The
*OPC? command returns ‘‘1’’ to the output buffer when the operation is complete.
The *OPC command sets the ‘‘OPC’’ bit (bit 0) in the Standard Event register
when the operation is complete.
Immediate Triggering
To select the immediate trigger source, send the following command.
TRIG:SOUR IMM
•When the IMMediate is selected as a trigger source, an INITiate command
immediately transfers the VOLT:TRIG or CURR:TRIG value to VOLT or CURR
value. Any delay is ignored.

Table of Contents

Other manuals for Agilent Technologies E3633A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3633A and is the answer not in the manual?

Agilent Technologies E3633A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3633A
CategoryPower Supply
LanguageEnglish

Related product manuals