Chapter 4 4-49
Troubleshooting
Greater than 1000: Personality Specific Error Messages
10161 Lower Custom Mask is Invalid!
The user-specified lower custom mask cannot be
resolved into a limit line.
10231 Measurement failed for unknown reasons.
Check instrument settings and restart measurement.
10154 Measurement not defined for Out of Band.
User has changed to an out-of-band frequency range.
The band measurement only operates in the selected
band.
10227 Measurement suspended until carrier is turned
off.
The receive channel power and the receive spur
measurements are specified with the attenuation set to
0 dB. To prevent overload, the frequency spectrum of
interest is monitored for signal levels which exceed a
specified threshold before setting the attenuator to
0 dB. If a carrier is found, this message is displayed and
the completion of the measurement will not occur until
the carrier is removed. The carrier check may be turned
off using the properties form under the front-panel
Mode Setup key. You may also change the signal
threshold which determines a carrier on the properties
form.
10155 No Fast ADC hardware installed. Meas
unavailable.
The analyzer cannot use sweep times of less than 5
msec when (Option B7D or Option AYX) is not
installed. Therefore the measurement will not be
executed.
10147 Opt B7D bootrom requires upgrade.
The (Option B7D) bootrom revision is not supported by
the currently loaded personality version. Refer to the
user’s guide for the personality in use.

Table of Contents

Other manuals for Agilent Technologies E4403B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4403B and is the answer not in the manual?

Agilent Technologies E4403B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4403B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals