Chapter 4 4-53
Troubleshooting
Greater than 1000: Personality Specific Error Messages
10170 The Cable Fault Measurement is active. Mode
Setup is disabled.
Mode setup is not available in the cable fault utility.
10177 There are no spurs to inspect.
The user has attempted to switch the
Inspect Spur
softkey to the On position after the measurement has
finished, but found no spurs.
10229 The regression portion failed.
This message occurs when (Option B7D) is not
functioning properly. Demodulation measurements
(modulation accuracy and code domain) might fail as a
result of this error.
10157 Tracking Generator hardware is not present.
Meas unavailable.
The measurement requires a built-in tracking
generator.
Unable to uninstall personality, file not
deletable.
This message occurs when you try to delete a
personality which has been marked as non-deletable.
The personality is marked non-deletable at the factory.
Contact your nearest service center for further
problems.
10144 Unknown demod status.
Demodulation is in an unknown state. Press
Preset. If
the error persists, contact your service center.
10160 Upper Custom Mask is Invalid!
The user-specified upper custom mask cannot be
resolved into a limit line. The format is incorrect.