4-54 Chapter4
Troubleshooting
Greater than 1000: Personality Specific Error Messages
10138 Valid GSM burst not found in frame (Burst
Type).
No active GSM bursts that match the selected Burst
Type have been detected in the RF input signal. The
search was performed over the complete GSM frame.
10140 Valid GSM burst not found in frame (Ref
Burst).
No active GSM bursts that match the selected Burst
Type have been detected in the RF input signal. The
search was performed using the Ref Burst type setting
over the complete GSM frame.
10139 Valid GSM burst not found in specified
timeslot (Burst Type).
No active GSM bursts that match the selected Burst
Type have been detected in the RF input signal. The
search was only performed over the specified timeslot
setting.