Chapter 11 313
Assembly Replacement Procedures
Top Brace
Replacement
1. Carefully position the top brace on the deck. There is an alignment
pin on the mid web/fan assembly that should match up with the
alignment hole on the top brace. Make sure that no coaxial cables
will get pinched underneath the brace.
2. Using the T-10 driver, replace the top screws first, and then the side
screws after the top screws are tightened. Torque to 9 inch pounds.
3. Replace the outer case. Refer to the “Instrument Outer Case”
replacement procedure.