Chapter 1 51
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
Questionable Calibration Skipped Negative Transition This command determines which bits
in the Questionable Calibration Skipped Condition register will set the corresponding bit
in the Questionable Calibration Skipped Event register when the condition register bit has
a negative transition (1 to 0). The variable <integer> is the sum of the decimal values of
the bits that you want to enable.
Questionable Calibration Skipped Positive Transition This command determines which bits
in the Questionable Calibration Skipped Condition register will set the corresponding bit
in the Questionable Calibration Skipped Event register when the condition register bit has
a positive transition (0 to 1). The variable <integer> is the sum of the decimal values of the
bits that you want to enable.
Questionable Calibration Extended Failure Register
Questionable Calibration Extended Failure Condition This query returns the decimal value
of the sum of the bits in the Questionable Calibration Extended Failure Condition register.
Mode All
Remote Command :STATus:QUEStionable:CALibration:SKIPped:NTRansiti
on <integer>
:STATus:QUEStionable:CALibration:SKIPped:NTRansiti
on?
Example STAT:QUES:CAL:SKIP:NTR 1 Align RF skipped is not
required.
Preset 0
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All
Remote Command :STATus:QUEStionable:CALibration:SKIPped:PTRansiti
on <integer>
:STATus:QUEStionable:CALibration:SKIPped:PTRansiti
on?
Example STAT:QUES:CAL:SKIP:PTR 1 Align RF skipped is required.
Preset 32767
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command