6 Operating in CI Mode
General Guidelines
144 5977B Series MSD Operating Manual
General Guidelines
Always use the highest purity methane (and other reagent gases, if
applicable). Methane must be at least 99.9995% pure.
Always verify the MSD is performing well in EI mode before switching to CI.
(See “To Verify EI System Performance” on page 131.)
Ensure the CI source and GC/MSD interface tip seal are installed.
Ensure the reagent gas plumbing has no air leaks. This is determined in PCI
mode, checking for m/z 32 after the methane pretune.
Ensure the reagent gas inlet line is equipped with gas purifiers (not applicable
for ammonia).