EasyManuals Logo
Home>Agilent Technologies>Water Pump>G7077B

Agilent Technologies G7077B User Manual

Agilent Technologies G7077B
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #42 background imageLoading...
Page #42 background image
2 Installing 8890 GC Columns
To Install a Capillary Column in the GC/MSD Interface Using the Self-Tightening Column Nut
42 5977B Series MSD Operating Manual
To Install a Capillary Column in the GC/MSD
Interface Using the Self-Tightening Column
Nut
This procedure is for the installation of a capillary column directly into the
analyzer using the Agilent recommended self-tightening column nut.
Materials needed
Interface tip seal (G3870-20542) (See Figure 6 on page 44.) (Not used with EI
SS or Inert source)
Knurled tip seal retainer (G3870-20547) (Not used with EI SS or Inert source)
Tip seal spring (G7005-20024) (Not used with EI SS or Inert source)
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
Flashlight
Magnifying loupe
Gloves, clean
Large (8650-0030)
Small (8650-0029)
Self Tightening column nut for GC\MSD interface (5190-5233)
Ferrules, Vespel
0.27 mm id, for 0.10 mm id columns (5062-3518)
0.37 mm id, for 0.20 mm id columns (5062-3516)
0.40 mm id, for 0.25 mm id columns (5181-3323)
0.5 mm id, for 0.32 mm id columns (5062-3514)
0.8 mm id, for 0.53 mm id columns (5062-3512)
Septum (may be old, used inlet septum)
Safety glasses

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7077B and is the answer not in the manual?

Agilent Technologies G7077B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7077B
CategoryWater Pump
LanguageEnglish

Related product manuals