EasyManuals Logo
Home>Agilent Technologies>Water Pump>G7077B

Agilent Technologies G7077B User Manual

Agilent Technologies G7077B
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #184 background imageLoading...
Page #184 background image
7 General Maintenance
To Clean the EI HES
184 5977B Series MSD Operating Manual
To Clean the EI HES
Materials needed
Abrasive paper (5061-5896)
Alumina abrasive powder (393706201)
Aluminum foil, clean
Cloths, clean (05980-60051)
Cotton swabs (5080-5400)
Glass beakers, 500 mL
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Solvents
Methanol (reagent-grade)
Methylene chloride (reagent-grade)
Acetone (reagent-grade)
Ultrasonic bath
Procedure
1 Disassemble the EI HES. (See “To Disassemble the EI HES” on page 181.)
2 Collect the following parts to be cleaned: (See Figure 58 on page 185.)
Filament mount
Source mount (does not get abrasively or ultrasonically cleaned)
Repeller
Source body
Extractor lens (5)
Post extractor lens 1 (4)
Post extractor lens 2 (3)
Ion focus lens (2)
Entrance lens (1)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7077B and is the answer not in the manual?

Agilent Technologies G7077B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7077B
CategoryWater Pump
LanguageEnglish

Related product manuals