3 Installing Intuvo 9000 GC Columns
To Install the GC/MSD Interface Tip Seal
74 5977B Series MSD Operating Manual
To Install the GC/MSD Interface Tip Seal
Materials needed
Interface tip seal (G3870-20542)
Tip seal spring (G7005-20024)
Knurled tip seal retainer (G3870-20547)
The interface tip seal must be in place for the CI source, the EI XTR source,
and HES. The interface tip seal, spring, and knurled tip seal retainer are not
used with EI SS or Inert source.
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Procedure
1 Verify that the CI source, EI XTR source, or HES is installed. The tip seal,
spring, and knurled tip seal retainer must not be installed when an EI SS or
Inert source is installed. (See Figure 17 on page 75.)
2 Remove the interface tip seal, spring, and knurled nut retainer from the ion
source storage box. In this order, slide the spring, tip seal, and knurled tip seal
retainer over the column sleeve.
CAUTION
Electrostatic discharges to analyzer components are conducted to the side
board where they can damage sensitive components. Wear a grounded
antistatic wrist strap, and take other antistatic precautions before you open
the analyzer chamber.
CAUTION
Always wear clean gloves while handling any parts that go inside the GC or the
analyzer chambers.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7077B and is the answer not in the manual?

Agilent Technologies G7077B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7077B
CategoryWater Pump
LanguageEnglish

Related product manuals