8CI Maintenance
To Remove the EI HES Radiator
224 5977B Series MSD Operating Manual
To Remove the EI HES Radiator
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Tweezers (8710-2460)
T10 Torx screwdriver (5182-3466)
Procedure
1 Remove the EI HES. (See“To Remove the EI HES” on page 178.)
2 Disconnect the wires from the EI HES radiator. (See“To Connect/Disconnect
Wiring to the EI HES Radiator” on page 225.)
3 Use a T10 Torx screwdriver to loosen the 2 screws that secure the radiator to
the analyzer, and place the radiator in its storage container.
WARNING
The analyzers, GC/MSD interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
Make sure you use an antistatic wrist strap and take other antistatic
precautions before touching analyzer components.
When disconnecting leads, pull on the connectors, not on the wires

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7077B and is the answer not in the manual?

Agilent Technologies G7077B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7077B
CategoryWater Pump
LanguageEnglish

Related product manuals