8CI Maintenance
To Disassemble the CI Source
238 5977B Series MSD Operating Manual
To Disassemble the CI Source
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Hex ball driver, 1.5 mm (8710-1570)
Hex ball driver, 2.0 mm (8710-1804)
Wrench, open-end, 10 mm (8710-2353)
Nut driver, 5.5 mm (8710-1220)
Tweezers (8710-2460)
Procedure
Refer to the exploded parts view Figure 78 and CI source parts list Table 24 on
page 239 while using this procedure.
1 Remove the CI source. (See“To Remove the CI Source” on page 235.)
2 Remove the filaments. (See“To Remove the CI Source Filament” on
page 247.)
3 Separate the source heater assembly from the source body. The source
heater assembly includes the source heater, repeller, and related parts.
4 Disassemble the repeller assembly by removing the ceramic insulator from
the repeller.
5 Remove the setscrew securing the lenses to the source body.
6 Pull the lenses out of the source body and separate the lens insulator, ion
focus lens, drawout cylinder, drawout lens, and entrance lens.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7077B and is the answer not in the manual?

Agilent Technologies G7077B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7077B
CategoryWater Pump
LanguageEnglish

Related product manuals