4 Installing 7890 GC Columns
To Condition a Capillary Column
5977B Series MSD Operating Manual 83
To Condition a Capillary Column
Materials needed
Carrier gas, (99.9995% pure or better)
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
Procedure
1 Install the column in the GC inlet. (See“To Install a Capillary Column in a
Split/Splitless Inlet” on page 80.)
2 Set a minimum velocity of 30 cm/s, or as recommended by the column
manufacturer. Allow the carrier gas to flow through the column at room
temperature for 15 to 30 minutes to remove air.
3 Program the oven from room temperature to the maximum temperature
limit for the column.
4 Increase the temperature at a rate of 10 to 15 °C/min.
5 Hold at the maximum temperature for 30 minutes.
6 Set the GC oven temperature to 30 °C, and wait for the GC to become ready.
7 Attach the column to the GC/MSD interface. (See“To Install a Capillary
Column in the GC/MSD Interface Using the Self-Tightening Column Nut” on
page 84.)
WARNING
Do not condition your capillary column with hydrogen. Hydrogen accumulation
in the GC oven can result in an explosion. If you plan to use hydrogen as your
carrier gas, first condition the column with ultrapure (99.999% or better) inert
gas such as helium, nitrogen, or argon.
WARNING
The GC operates under high temperatures. Do not touch any GC parts unless
you are certain they are cool.
CAUTION
Never exceed the maximum column temperature, either in the GC/MSD
interface, the GC oven, or the inlet.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7077B and is the answer not in the manual?

Agilent Technologies G7077B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7077B
CategoryWater Pump
LanguageEnglish

Related product manuals