6 Operating in CI Mode
To Monitor CI Mode High Vacuum Pressure
166 5977B Series MSD Operating Manual
Familiarize yourself with the measurements on your system under operating
conditions and watch for changes that may indicate a vacuum or gas flow
problem. Measurements will vary by as much as 30% from one MSD and gauge
controller to the next.
Table 16 MFC settings and typical pressure readings
MFC Pressure (Torr)
(%) Methane Ammonia
EI/PCI/NCI MSD
(Turbo pump)
EI/PCI/NCI MSD
(Turbo pump)
10 5.5
×10
–5
5.0 ×10
–5
15 8.0 ×10
–5
7.0 ×10
–5
20 1.0 ×10
–4
8.5 ×10
–5
25 1.2 ×10
–4
1.0 ×10
–4
30 1.5 ×1
–4
1.2 ×10
–4
35 2.0 ×10
–4
1.5 ×10
–4
40 2.5 ×10
–4
2.0 ×10
–4