3 Installing Intuvo 9000 GC Columns
Handling the Intuvo 9000GC Column and Bus Components
58 5977B Series MSD Operating Manual
Handling the Intuvo 9000GC Column and Bus
Components
The Agilent Intuvo 9000 Gas Chromatograph (Intuvo 9000 GC) does not use
traditional ferrules and nuts for most column and flow path seals. In a traditional
gas chromatography connection, the seal is made by deforming a soft ferrule
around the periphery of a column or tube, with a second seal made between the
ferrule and the fitting. Instead, the Intuvo 9000 GC click and run connections use
a sealing system based on contact between flat surfaces. Compared to
traditional ferrule seals, these connections are leak-free and easy to make.
When making these seals, follow a few simple guidelines:
Do not touch the click and run sealing surfaces with bare skin or dirty gloves.
Skin oils and dirt can contaminate the flow path surfaces.
Use only the provided Intuvo 9000 GC torque driver to tighten Intuvo
compression bolts.
Avoid scratching or deforming click and run sealing surfaces.
If you need to clean a sealing surface, use clean, compressed air.
Use a new gasket each time you install a column or Intuvo chip.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7077B and is the answer not in the manual?

Agilent Technologies G7077B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7077B
CategoryWater Pump
LanguageEnglish

Related product manuals