EasyManuals Logo
Home>Agilent Technologies>Water Pump>G7077B

Agilent Technologies G7077B User Manual

Agilent Technologies G7077B
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #88 background imageLoading...
Page #88 background image
4 Installing 7890 GC Columns
To Install a Capillary Column in the GC/MSD Interface Using the Self-Tightening Column Nut
88 5977B Series MSD Operating Manual
15 Gently check the alignment of the ion source and the interface tip seal.
When the ion source is aligned correctly, the analyzer chamber can be
closed all the way with no resistance except the spring tension from the
interface tip seal.
16 You can align the ion source and interface tip seal by wiggling the side plate
on its hinge. If the door still will not close, contact your Agilent Technologies
service representative.
17 Close the analyzer chamber door. (See “To Close the Analyzer Chamber” on
page 220.)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7077B and is the answer not in the manual?

Agilent Technologies G7077B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7077B
CategoryWater Pump
LanguageEnglish

Related product manuals