Splitless Inlet Troubleshooting
Most problems encountered with a splitless injection are related to:
Incorrect purge time
Degradation
Improper focusing
Inappropriate column temperature
Flashback
You can also improve the reproducibility and linearity of peak areas and avoid backflash
by matching:
Inlet temperature
Liner volume
Injection volume
Decomposition
Loss of peak area or generation of new peaks, can sometimes be dramatically reduced by
changing liner type or by deactivating the liner and inlet with silanizing reagents.
Removing or reducing the amount of liner packing can also decrease inlet activity.
Column Troubleshooting
Normal Peaks Tailing Solvent Peaks Wrong Peak Ratios
Correct column positioning in
both injection port and FID
Column positioned incorrectly
in the injection port of
possible ferrule particle in
the carrier gas flow path
Column positioned in the inlet
(either too far or not far
enough; verify 4-6 mm
installation distance)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies gc/ms series and is the answer not in the manual?

Agilent Technologies gc/ms series Specifications

General IconGeneral
BrandAgilent Technologies
Modelgc/ms series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals