The GTA Viewing/Fume Extraction Accessory (Optional)
Agilent Graphite Tube Atomizer GTA 120 User’s Guide 47
Figure 15. Note the correct location of the edge of the flame shield in relation to
the spring clip on the accessory
Figure 16. The canoe-clips hold the accessory in place firmly against the flame
shield
4 The base of the accessory should sit 2–3 mm above the
magnet/workhead. Loosen the mounting bracket screws and set
the accessory to the correct height. Tighten the screw firmly.
5 Check that the accessory is firmly in place and cannot move from
side to side. The curve at the bottom of the accessory should be
centered over the sample port. Refer to Figure 17.