The GTA Viewing/Fume Extraction Accessory (Optional)
48 Agilent Graphite Tube Atomizer GTA 120 User’s Guide
Figure 17. The rim of the accessory should be centered over the sample port
NOTE
To adjust the position of the accessory, tilt the base forwards and lift it up
slightly to enable the accessory to move freely. Move the accessory to the
desired position and then press it down to lock into place.
6 Plug the accessory into the accessory port on the instrument
using the 9 way D-type connector fitted to the accessory. The
accessory port is located on the rear right hand side of the
instrument.
Installing the Extraction System to the Laboratory Exhaust System
The GTA viewing/fume extraction accessory is connected to the
extraction duct using an adaptor plate. Two plates are supplied with
the accessory: use the longer one if you have a 150 mm (5 in.) duct
and the smaller one if you have a 125 mm (5 in.) duct. If your duct is
a different size, contact your local Agilent field service engineer.
Leave a 2–3 mm gap between the
furnace and the rim of the
accessory.