Performance Tests and Calibration 5
U1241B/U1242B User’s and Service Guide 41
Test Considerations
Error may be induced by AC signals that present on the input leads. Long 
test leads can also act as an antenna causing pick- up of AC signals.
For optimum performance, all procedures should comply with the 
following recommendations:
• Ensure that the calibration ambient temperature is stable and between 
18 °C and 28 °C. Ideally the calibration should be performed at 23 °C 
± 2 °C.
• Ensure that the ambient relative humidity (RH) is less than 80%.
• Ensure that the low battery indicator does not appear during the 
performance verification test. If it does, replace the batteries to avoid 
any inaccurate reading.
• During temperature performance verification test, ensure that the 
multimeter has been switched ON and placed in the test environment 
for at least one hour with J/K- type thermo- couple connected between 
the multimeter and calibration source.
• Allow one minute warm- up period with a Shorting Plug connected to 
the V and COM input terminals.
• Use shielded twisted pair PTFE- insulated cables to reduce settling and 
noise errors. Keep the input cables as short as possible.
• Connect the input cable shields to earth ground. Except where noted in 
the procedures, connect the calibrator LO source to earth ground at the 
calibrator. It is important that the LO to earth ground connection be 
made at only one place in the circuit to avoid ground loops.
During DC voltage, DC current, and resistance gain verification 
measurements, ensure that the calibrator's "0" output is correct. It is 
recommended to set the offset for each range of the measuring function 
being verified.