7Performance Tests
Agilent U1602B/U1604B User’s and Service Guide 139
4 Press Autoscale on the instrument.
5 Select Measure mode, then press F1 to enable the auto measurement
selection for Measure 1. Turn rotary switch to select peak- to- peak
voltage (Pk- Pk) measurement from the list of auto measurement.
Wait for a few seconds for the measurement to settle.
6 View the Pk- Pk reading at the measurement line display. Record the
reading: Vp- p = _______ V. (refer to “Agilent U1602B/U1604B Handheld
Digital Oscilloscope Test Record”)
7 Change the frequency of the signal generator to the value shown below
for your instrument.
Table 7 - 4 Signal generator frequency setting
8 Change the time base to 50 ns/div.
9 Wait for a few seconds for the measurement to settle.
10 View the Pk- Pk reading at the measurement line display.
11 Record the reading: Vp- p = ______ mV.
12 Calculate the bandwidth response (dB) using the following formula:
20log
10
[Reading from step #11 / Reading from step #6]
13 The result from step 12 should be greater than or equal to –3.0 dB.
14 Move cable from channel 1 to channel 2 input and repeat the same
verification procedures for channel 2.
Selected Channel U1602B U1604B
Channel 1 20 MHz 40 MHz
Channel 2 20 MHz 40 MHz

Table of Contents

Other manuals for Agilent Technologies U1604B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1604B and is the answer not in the manual?

Agilent Technologies U1604B Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1604B
CategoryTest Equipment
LanguageEnglish

Related product manuals