9 Characteristics and Specifications  
Agilent U1602B/U1604B User’s and Service Guide  173
[1]  For temperature between 0 
o
C to 18 
o
C and 28 
o
C  to 50 
o
C , add 0.1% of reading + 0.02% of range 
for every degree Celsius.
DIGITAL MULTIMETER SPECIFICATIONS
[1]
 ± (% of reading + % of range)
Function Range Frequency, Test Current or 
Burden Voltage
1 year Tcal ± 5 
o
C
DC Voltage 600.0 mV 0.3 + 0.08
6.000 V 0.3 + 0.08
60.00 V 0.3 + 0.08
600.0 V 0.3 + 0.08
AC Voltage 600.0 mV – 600.0 V 50 Hz – 1 kHz 1.0 + 0.2
1 kHz – 30 kHz 3.0 + 0.2
AC+DC Voltage 6.0000 V – 600.0 V 50 Hz – 1 kHz 1.0 + 0.2
1 kHz – 30 kHz 3.0 + 0.2
Resistance 600.0 W 0.5 + 0.2 
6.000 kW 0.5 + 0.2 
60.00 kW 0.5 + 0.2 
600.0 kW 0.5 + 0.2 
6.000 MW 0.5 + 0.2 
60.00 MW 1.0 + 0.2 
Capacitance 60.00 nF 2.0 + 0.2
600.0 nF 2.0 + 0.2
6000 nF 2.0 + 0.2
60.00 μF2.0 + 0.2
300.0 μF2.0 + 0.2
Diode 1.000 V 0.5 mA 2.0 + 0.08