28 U1731B/U1732B User’s and Service Guide
3 Service and Maintenance
Table 3 - 3 Inductance ranges for functional validation
Inductance (Series Mode), Test Frequency: 100 Hz, 120 Hz, 1000 Hz, or 10 kHz
Recommended Equipment: GR1491 Precision
Decade Inductor
Range (H) Test Value Used
200 m 100 m
20 m 10 m
2000 µ
*
* Does not support test frequency of 100 Hz and 120
Hz
1000 µ