Soft Start
17
TwisTorr 74 FS User Manual / 87-901-053-01 307/342
Soft Start
To start the pump after a protracted stop of the TwisTorr 74 FS, it’s
recommended to enable the “Soft Start” function; this allows a better
grease distribution in the bearings.
The “Soft Start” mode is disabled by default.
To enable the “Soft Start” mode it must be activated by the suitable
software or Remote I/O command (see Controller Manual).
“Soft Start” frequency steps are the following: 450Hz, 1100Hz, 1167Hz
or more details see Controller Manual.