17 Technical Information
Connection of the High Vacuum Flange
324/342 TwisTorr 74 FS User Manual / 87-901-053-01
Connection of the High Vacuum Flange
In the case the rotor is suddenly blocked, the torque arising from the
system and the high vacuum flange must be absorbed. Only the
components listed in the following can be used to fasten the pump to
the high vacuum flange.
WARNING!
It is mandatory to connect the pump in such a way to withstand the torque
specified in the "INSTRUCTION FOR USE" section. Specifically pay attention
to clamp design, material of clamps and bolts and bolt fixing torque.
WARNING!
If the pump is fastened on a vacuum chamber with different flange Agilent
there exists the chance of twisting or tearing-off in case the rotor is suddenly
blocked. The combination of different flange types is not allowed in any case.
Agilent Vacuum will not accept any liability for all damages resulting from
not allowed fastening.
Installation of ISO-K Flange
For ISO-K flange connections, fix the two flanges with the clamps as
shown in the following figure (a protective screen can optionally be
used).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies X3502-64170 and is the answer not in the manual?

Agilent Technologies X3502-64170 Specifications

General IconGeneral
BrandAgilent Technologies
ModelX3502-64170
CategoryWater Pump
LanguageEnglish

Related product manuals