Problem Probable cause Solution
Ghosting or double exposure
on X-ray image
PSP exposed twice Only expose the PSP once
PSP not su󰀩ciently erased Check the erasure unit for proper function.
Inform a service technician, if the problem per-
sists.
X-ray image mirrored in one
corner
PSP bent during X-ray exposure Do not bend the PSP
Shadow on the X-ray image PSP removed from the barrier
envelope before scanning
Do not handle PSPs without a barrier envelope.
Store the PSP in a barrier envelope.
X-ray image cut o󰀨, part
missing
A metal part of the X-ray tube is
in front of the X-ray beam
Recording an X-ray image, make sure there are
no metal parts between the X-ray tube and the
patient.
Check X-ray tube.
Faulty edge masking in imaging
software
Deactivate edge masking.
Software unable to combine
the data to make a complete
image
The X-ray dose on the PSP was
insu󰀩cient
iiIncrease X-ray dose.
Amplication (HV value) is set
too low in the software
Increase amplication (HV value).
Unsuitable scanning mode
selected
Select a suitable scanning mode.
The setting for the threshold
value is too high
Reduce the threshold value