5
EN
Mode Maximum Input
A~, A 10 A, 1000V fast acting
fuse (max. 30seconds
every 15minutes)
mA~, mA 800 mA, 1000V fast
acting fuse
V~, V 1000 V~ rms/
Frequency, Resistance,
Capacitance, Diode
Test, Continuity
250 V~ rms /
Surge Protection: 8 kV peak per IEC 61010
• Do not measure current on a CAT III circuitry
whose voltage exceeds 1000 V.
• Do not measure current on a CAT IV circuitry
whose voltage exceeds 600 V.
• When measuring volts, do not switch to
current/resistance modes.
• Set the function switch to the appropriate
position before measuring.
• Set the function switch to the OFF position
when not in use.
• When changing ranges always disconnect the
connector leads from the circuit under test.
• Inspect the condition of the connector leads
and the product itself for any damage before
operation. Repair or replace any damage
before use.