4-10 5000 Multi Point Moisture Analyzer
Alarm_1_Hi 60
Alarm_1_Low 0
Alarm_2_Hi 15
Alarm_2_Low 0
5) Set timers to achieve sample switching frequency of 10 minutes including equilibration
periods between sample analysis and to set duration of calibration cycle to 10 minutes. Press
the test/config key; select Configure Menu-timing-Sample_Cycles; set the following timers to
8 cycles.
SamplCycl_1
SamplCycl_2
SamplCycl_5
Return to Timing menu; select Equil_Cycles; set the following timers to 2 cycles.
EquilCycl_1
EquilCycl_2
EquilCycl_5
AMETEK 5000 Controller
with Multi-Point
Default System Security Passwords
Level 1: 1 1 1 1 [ENTER]
Level 2: 2 2 2 2 [ENTER]
Level 3: .3 - 18 [ENTER]