110-512 BEL/EN
28
Setting the Low Limit Temperature
This sub-menu allows you to limit how low the temperature set point may be set. It also serves as a low
limit safety, alerting you if bath temperature falls below the low limit temperature setting. The Low Limit
value may be set from -52° to +20°C / -65° to +65°F; the factory default is -52°C.
To avoid any unwanted alarms or shutdown during regular operation, the Low Limit value should be set at
least 5° lower than the selected control temperature.
To Access: Press the
key until LOW LIMIT appears on the display.
To Change: Touch the arrows until the desired low limit temperature is displayed.
To Accept: Press
,
, ,
or allow the LCD to return to the main operational
display (approximately 10 seconds).
Touch and hold the up/down
arrows to make large
changes
Touch and release to make
minor changes