www.amico.com 19
The manual emergency lowering and raising feature should only be used if all other
controls fail. A proper safety ladder or stool may be required to remove the plug from
the cover. Remove the round plug from the lift cover and use a 3/16" or 2.5 mm Allen
key to rotate the motor in the up or down direction.
!
CAUTION: DO NOT attempt to use the lift while using manual lowering.
Overspeed Cam
The Overspeed Cam brake is made of a metal bar fixed to the drum. In case of gear or motor breakage, the centrifugal
force created will block the bar against the frame.
Cleaning and Disinfection
The exterior of the GoLift should only be cleaned, disinfected using the recommended cleaning agents shown below.
Damp a cloth with the cleaning agent and wipe down entire exterior of the lift and Carry Bar. Other chemicals and/or
liquids not listed should not be used to clean and disinfect this lift.
!
CAUTION: Take great care to ensure that no liquids get inside the Amico GoLift. The lift is not drip proof or
water tight. Failure to protect the lift from liquids may result in damage to the lift and may cause personal
injury.
RECOMMENDED CLEANING AGENTS
CaviWipes
Clorox Healthcare Professional Disinfecting Bleach Wipes
Dispatch Hospital Cleaner Disinfectant Towels With Bleach
Oxivir Tb Disinfectant Wipes
Sani-Cloth Super Germicidal Disposable Wipes
tb Minuteman NEX GEN
Virocidin-X
Virox-5
Virox Accel Tb