28 Amico Mobility Solutions Corporation
Electro-Magnetic Compliance Data for Amico GoLift
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment specified below.
The customer or the user of the [EQUIPMENT or SYSTEM] should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic Environment –
Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV for power supply lines
±0.250 kV for input/output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the [EQUIPMENT or SYSTEM]
requires continued operation
during power mains interruptions,
it is recommended that the
[EQUIPMENT or SYSTEM] be
powered from an interruptible
power supply or a battery.
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
3 A / m Not Applicable
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.