Page 12 OM001-16 Otowave 102 Operating Manual
5.2. Ear tip(s)
These must be selected and fitted by a practitioner qualified to perform
tympanometric tests.
The probe tip must be fitted with a new ear tip before it
is presented to a patient’s ear canal. The ear tip must
be fitted completely to the probe tip and must not
occlude any of the four holes in the probe tip. The ear
tip size is chosen to suit the patient’s ear and provide
a comfortable pressure seal.
Refer to Sections 2.1 and 11.2 regarding these single-
use parts.
5.3. Performing a test
Having selected the required test settings a typical tympanogram
measurement and reflex tests are carried out as follows.
From the MAIN MENU select NEW TEST:
MAIN MENU
NEW TEST
VIEW THE LAST TEST
DAILY CHECK
Select
Select the ear(s) required for test:
SELECT EAR
BOTH
LEFT
RIGHT
Back  Select