Page 32 OM001-16 Otowave 102 Operating Manual
performed
centre.
WARNING! DEFAULTS RELOADED.
Default configuration settings reloaded.
Check before making new tests
This message should never be
seen. Check all the
CONFIGURATION settings
before taking measurements. If
the error persists, contact your
Amplivox service centre.
ERROR
Transfer failed
No device found or Link was unreliable
The Otowave was unable to
send data to the computer. See
Section 8 for details.
WITHDRAW PROBE
The probe has been moved
during measurement. Re-insert
the probe to repeat the test.
Volume outside range
WITHDRAW PROBE
The ear canal volume is above
the 5ml. This message also
occurs when the probe is not
properly inserted into the ear.
Blocked probe
WITHDRAW PROBE
The ear canal volume is below
0.1ml. This message also
occurs when the probe tip is
blocked. Check that the probe
is correctly inserted into the
ear. Check that the probe is not
blocked.
INSERT PROBE
The seal was lost. Reinsert the
probe to repeat the test.
14. Technical Specification
14.1. Performance
Tympanometry
Instrument type
Meatus compensated tympanometer
Analysis performed
Compliance peak level (in ml) &
pressure; Gradient (in daPa);
Ear Canal Volume (ECV) @ 200 daPa
Probe tone levels and accuracy
226Hz +/- 2%; 85dB SPL +/-2dB over
range 0.2ml to 5ml
Pressure levels and accuracy
+200daPa to -400daPa +/-10daPa or
+/-10% (whichever is larger) over range