Handle the probe and accessories with care. Do not
allow moisture, condensation, fluids or debris to enter
the probe.
11.3. Calibration and Repair of the Instrument
Amplivox recommends that the Otowave is calibrated annually. Please
contact Amplivox for details.
If the instrument is to be used at elevations above that specified in Section
2.1 re-calibration must be undertaken at the intended operating elevation.
The instrument should be returned to the manufacturer
for service & repair. There are no user-serviceable
parts within it.
Please use the original shipping carton and packaging to transport the
instrument. Place the instrument in a plastic bag before packing to prevent
dirt and dust getting into the probe. Do not return the batteries with the
instrument.