Command Directory 3-4 Commands
MG369xC GPIB PM PN: 10370-10374 Rev. F 3-57
LIN
Description: Selects linear power level operation in mV.
In linear power level operation, power level entries and outputs are in mV and power
level sweeps are linear.
Functional
Category: POWER LEVELING
Related
Commands: LOG, PU0, PU1, PU2
Front Panel Key: Level | Level | Level Control > | Units > | mV
LIS
Description: Selects linear step sweep.
LIS, by itself, does not provide a swept-frequency output. It only determines that the
swept-frequency output will be a linear step sweep. However, if a frequency sweep is
being output by the instrument when LIS is received, the sweep will become a linear step
sweep. This is the default mode.
Functional
Category: SWEEP
Related
Commands: SWP, SSP, LGS, MAN
Example: SDT 3 MS; SNS 100 SPS; LIS; SF1; AUT
Initiates a continuous linear step frequency sweep between F1 and F2. The sweep
consists of 100 steps with a dwell-time-per-step of 3 ms.
Output Command: OSD returns the dwell time of the step sweep (in ms) to the controller. OSS returns the
number of steps of the step sweep to the controller. OST returns the sweep time value
(in ms) to the controller.
Front Panel Key: Frequency | Step Sweep | More > | Linear
LO0
Description: Turns off the level offset function.
Functional
Category: POWER LEVELING
Related
Commands: LO1, LOS
Front Panel Key: Level | Level | Offset to Off

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu MG369 C Series and is the answer not in the manual?

Anritsu MG369 C Series Specifications

General IconGeneral
BrandAnritsu
ModelMG369 C Series
CategoryPortable Generator
LanguageEnglish

Related product manuals