3-4 Commands Command Directory
3-56 PN: 10370-10374 Rev. F MG369xC GPIB PM
LGS
Description: Selects logarithmic step sweep.
LGS, by itself, does not provide a swept-frequency output. It only determines that the
swept-frequency output will be a logarithmic step sweep. However, if a frequency sweep
is being output by the instrument when LGS is received, the sweep will become a
logarithmic step sweep.
Functional
Category: SWEEP
Related
Commands: SWP, SSP, LIS, MAN
Example: SDT 3 MS; SNS 100 SPS; LGS; SF1; AUT
Initiates a continuous logarithmic step frequency sweep between F1 and F2. The sweep
consists of 100 steps with a dwell-time-per-step of 3 ms.
Output Command: OSD returns the dwell time of the step sweep (in ms) to the controller.
OSS returns the number of steps of the step sweep to the controller.
OST returns the sweep time value (in ms) to the controller.
Front Panel Key: Frequency | Step Sweep | More > | Log
LIB(xxxx)
Description: Sets the list start index to xxxx.
In list sweep mode, permits setting the list start index for a list sweep of the current list.
Functional
Category: LIST SWEEP
Value: xxxx: 4-digit integer between 0000 and 1999
Example: LST; ELN1; LIB1097; LIE1197; EXT; LDT 10 MS; TRG
Places the unit in list sweep mode and implements a list sweep from list 1 index 1097 to
index 1197 in single trigger mode with a 10 ms dwell-time-per-step, then triggers a single
sweep.
Related
Commands: LST, ELN(x), LIE(xxxx)
Front Panel Key: Frequency | List | Sweep > | Start Index
LIE(xxxx)
Description: Sets the list stop index to xxxx.
In list sweep mode, permits setting the list stop index for a list sweep of the current list.
Functional
Category: LIST SWEEP
Value: xxxx: 4-digit integer between 0000 and 1999
Example: LST; ELN1; LIB1097; LIE1197; EXT; LDT 10 MS; TRG
Places the unit in list sweep mode and implements a list sweep from list 1 index 1097 to
index 1197 in single trigger mode with a 10 ms dwell-time-per-step, then triggers a single
sweep.
Related
Commands: LST, ELN(x), LIB(xxxx)
Front Panel Key: Frequency | List | Sweep > | Stop Index

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu MG369 C Series and is the answer not in the manual?

Anritsu MG369 C Series Specifications

General IconGeneral
BrandAnritsu
ModelMG369 C Series
CategoryPortable Generator
LanguageEnglish

Related product manuals