Command Directory 3-4 Commands
MG369xC GPIB PM PN: 10370-10374 Rev. F 3-123
XL0 <value> <unit>
Description: Opens the L0 parameter.
Permits setting the L0 parameter. The power level setting is determined by the power
level range of the particular MG369xC model.
Functional
Category: PARAMETER ENTRY
Value: Power level range of the MG369xC model
Units: DM (log), VT (linear)
Output Command: OL0 returns the L0 power level (in dBm when in log mode; in mV when in linear mode)
to the controller.
Related
Commands: LOG, LIN, PU0, PU1, PU2
Front Panel Key: Level | Level | Level Control > | highlight L0, then press Edit Selected
XL1 <value> <unit>
Description: Opens the L1 parameter.
Permits setting the L1 parameter. The power level setting is determined by the power
level range of the particular MG369xC model.
Functional
Category: PARAMETER ENTRY
Value: Power level range of the MG369xC model
Units: DM (log), VT (linear)
Output Command: OL1 returns the L1 power level (in dBm when in log mode; in mV when in linear mode)
to the controller.
Related
Commands: LOG, LIN, PU0, PU1, PU2
Front Panel Key: Level | Level | Level Control > | highlight L1, then press Edit Selected
XL2 <value> <unit>
Description: Opens the L2 parameter.
Permits setting the L2 parameter. The power level setting is determined by the power
level range of the particular MG369xC model.
Functional
Category: PARAMETER ENTRY
Value: Power level range of the MG369xC model
Units: DM (log), VT (linear)
Output Command: OL2 returns the L2 power level (in dBm when in log mode; in mV when in linear mode)
to the controller.
Related
Commands: LOG, LIN, PU0, PU1, PU2
Front Panel Key: Level | Level | Level Control > | highlight L2, then press Edit Selected