5.5. Micromotor
Coupling the handpieces and changing the chuck.

Use.
WARNING!
Also read the instructions for use of the various motors.
Operating time: work 5 min., rest 5 min.


f 
e 
Adjusting the micromotors operating speed.
.


NOTE: the data set are automatically saved.


Selecting the direction of rotation.


󰀨

WARNING!
When the micromotor is taken out, 3 BEEPS sound to signal counter-
clockwise rotation.
NOTE: when the rheostat switch is on, the micromotor cannot
reverse.


󰀨
󰀨

NOTE: the bre optics shut o when the micromotor is not used for
30 seconds (rheostat switch o)
Cleaning and care.


WARNING!
• Do not soak the syringe in liquid disinfectants or detergents.
• Products not recommended: harsh products and/or products containing acetone, chlorine and sodium hypochlorites.
Sterilization.

WARNING!
Carefully read the operating instructions supplied with the instrument before attempting to sterilize.
Safety guidelines
WARNING!
Never put the contra angle on the micromotor while it is running.
The chuck release button must be held down during operation!
Friction between the button and micromotor rotor overheats the head and may cause burns.
The patient’s internal tissues (tongue, cheeks, lips, etc...) must be protected against contact with the button by using suitable instruments
(mirror, etc...).
• The chucks and various instruments attached to the handpieces must comply to the standard ISO 10993.