6.6. Hydraulic saliva ejector
ThehydraulicsalivaejectorstartsrunningWhenthetubeisremovedfrom
the support.
Cleaning after each use.
Aspirateabout½litreofSTER3PLUS(CEFLAs.c.)dilutedina6%solution
(equivalentto60mlofproductin1litreofwater).
Cleaning the saliva ejector lter.
Thisoperationmustbecarriedoutattheendofeachworkday.
WARNING!
Put on gloves before attempting to perform this operation!
• Aspirateabout½litreofSTER3PLUSdilutedina6%solution(equivalent
to60mlofproductin1litreofwater).
• In order to prevent possible dripping of liquids and secretions from the
lter(b)tobeextracted,aspirateonlyairforabout5seconds.
•Takeothecap(a ) by turning and twisting at the same time.
•Removethelter(b ).
•Clean/replacethelter(code97290060).
•Putthelterandcapbackinplace.
NOTE: to prevent liquids and matter from dripping from the lter
taken out, suck air only for approximately 5 seconds before performing
these operations.
Routine maintenance.
Lubricate the o-rings ( c ) with S1 – Protection for o-rings lubricant.
7. Hydrogroup operation
7.1. Fill cup and bowl
Thebowlcanbeturned160°onthehydrogroupbyhand.
Thewatertocupspoutcanberemovedtoeasecleaning.
Control buttons.
Water to bowl.
Water to cup.
Bowlushingautomaticallystopsafter30seconds.
Bowlushingautomaticallystartsinthefollowingcases:
• When the "rinse position” button is pressed;
• When the "water to cup” button is pressed.
Hot water to cup.
Whenthisfunctionisprovided,hotwaterisalwaysdeliveredtothecup.
Adjusting the amount of water used to ll the cup.
•TurnontheoperatingunitholdingdowntheCupWaterDeliverybutton
on the assistant’s board.
NOTE: 2 BEEPs signal saving is in progress.
•Pressthewatertocupkeyoncetostartllingthecup.
•Oncethedesiredwaterlevelhasbeenreached,pressthewatertocup
button again making sure the BEEP stops:
Atthispoint,thenewwaterlevelinthecuphasbeensaved.