7.5. Automatic tool ushing cycle (FLUSHING)
Description of the system.
TheautomaticFLUSHINGcycleallowstocarryoutanautomaticushing
cycle to renew water present in the water ducts of the instruments on the
dentist’s and the assistant’s boards and the water-to-cup duct.
Flushingmaybecarriedoutwithmainswater,treatedwater(iftheWHE
system is present) or distilled water (if the SHS system is present)
Thecycletimeisabout2minutes.
WARNING!
It is advisable to carry out a FLUSHING cycle at the beginning of each
working day and between two patients.
FLUSHING cycle setting.
•IftheS.H.S.systemispresentandyouwanttoexecutetheushingcycle
withdistilledwater,checkthatthelever(b)isontheTANKposition.
NOTE: it is advisable to execute the ushing cycle with a full
tank ( a ).
• Press the button onthedentist’sconsoletoaccesstheFLUSHINGmode.
NOTE: Three animated symbols will be shown on the console
display to conrm that FLUSHING mode has been activated.
• Check that the spray taps ( d ) in the lower part of the dentist’s instrument
boardareopen(ifnot,noortoolittlewaterwillcomeout).
• Insert the special supplied container ( c ) under the cup spout.
• Position the container ( e ) for the instruments to be disinfected on the bowl.
• Insert the tubings of the instruments to be disinfected in the container.
WARNING!
For the syringe instrument you need to use the special adapter ( f )
and the heating system must be o.
The micromotor tubing must be completely inserted into the motor
body.
Execution of the FLUSHING cycle.
•Starttheushingcyclebypressingthebutton on the dentist’s console
a second time.
NOTE: the time remaining to the end of the ushing cycle is shown
on the console display.
•AttheendoftheFLUSHINGcycle(thevalue0isshownonthedisplay),
theLEDsoftheFIBREOPTICSandINVERSIONbuttonsstartashing.
•Atthispoint,puttheinstrumentsextractedbackintoplacetoreturntothe
working condition.
Interruption of the FLUSHING cycle.
Oncetheushingcyclehasbeenstarted,itCANNOTbeinterrupted.
Error signalling.
Signalling:2beepswhenyouattempttostarttheushingcycle.
Cause: No instrument extracted or instruments not equipped with water
ducts extracted.
Remedy:Checktheinstrumentsextractedandthenstarttheushingcycle.
NOTE: if the error message appears again, call Technical Service.