11. Dental operatory maintenance plan
WHEN PART HOW SEE PARAGRAPH
Before starting work.
Draincondensate. / See paragraph 9.2
CATTANIsurgicalseparator.
Insertinsideeachsuctiontubelteratablet
of VF CONTROL PLUS
See paragraph 9.5
After each treatment.
Contra angle handpiece. Sterilize or disinfect the outside.
See documentation enclosed
with handpiece
Turbine. Sterilize or disinfect the outside. See paragraph 5.4
Micromotor. Disinfectoutside. See paragraph 5.5
Scaler. Sterilize or disinfect outside. See paragraph5.6
Syringe. Sterilize or disinfect outside. See paragraph 5.3
Curing lamp. Sterilizebreoptic,disinfectoutside. See paragraph 5.7
C-U2. camera.
Disinfectoutside.
Do not use acids or harsh products.
See paragraph 5.8
Surgical suction tubes.
Aspirate about ½ litre of sanitizing solution
with each cannula.
Sterilize the cannula holder terminals.
See paragraph 9.4
When needed.
Bowl.
Cleanwitho-shelfdetergentsformulatedfor
ceramic materials
Do not use acids or harsh products.
See paragraph 7.1
CATTANIgravityseparator. Drainseparator’scontainer. See paragraph9.6
METASYSsurgicalseparator.
See documentation supplied with the
equipment.
/
DURRsurgicalseparator.
See documentation supplied with the
equipment.
/
Operatory light.
See documentation supplied with the
equipment.
/
Monitor with light pole.
See documentation supplied with the
equipment.
/
Removable instrument cords.
Clean with a suitable disinfectant carefully
following the directions for use provided by
the manufacturer. Spray the product on soft
disposable paper towels.
Do not use acids or harsh products.
See paragraph 5.
Coated surfaces and dental
chair upholstery.
Clean with a suitable disinfectant carefully
following the directions for use provided by
the manufacturer.
Do not use acids or harsh products.
See paragraph 1.4
At the end of
the work day.
Bowllter.
Cleanlterinrunningwater
Thecontentmustbedisposedofseparately
See paragraph 7.1.
Surgicalsuctionlter.
Checkthelterandreplaceitifthesuctionrate
isreduced(code97461845).
See paragraph 9.3.
Surgical suction tubes.
Aspirate about ½ litre of sanitizing solution with
each cannula and then dry.
Sterilize the cannula holder terminals.
See paragraph 9.4.
Hydraulicsalivaejector. Cleanthelterofthesalivaejectorterminal.
Seeparagraph6.6
Weekly.
CATTANIsurgicalseparator.
Cleantheseparator’scontainer, drain valve
and probes.
See paragraph 9.5.
Cannula holder terminals. Lubricate the O-rings. See paragraph 9.4.
Monthly. Turbinereturnairlter.
Check the lter and replace it if necessary
(code 97290014).
See paragraph 9.7.
Yearly. Dentalchair.
Contact the technical service department for
general inspection.
/