Changing bowl operation
Waterdeliverytothebowlcantakeplaceinanuntimedmanner(on/o
operation with the relative button).
Thismodicationismadebyturningontheoperatingunitholdingdown
theBowlWaterDeliverybuttonontheassistant’sboard.
A beep sounds to signal the selected operating mode:
• 1 BEEP: timed operation.
• 3 BEEPS: ON/OFF operation.
NOTE: timed operation is factory set.
Changing the water to bowl controller with water delivered to cup.
Thecontrollerthatdeliverswatertothebowlwhenthe"watertocup”button
is pressed can be enabled/disabled.
Thismodicationismadebyturningontheoperatingunitsimultaneously
holdingdowntheCupWaterDeliveryandBowlWaterDeliverybuttonson
the assistant’s board.
A beep sounds to signal the selected operating mode:
• 1 BEEP: controller enabled.
• 3 BEEPS: controller disabled.
NOTE: the controller is enabled by default.
Water to bowl controller modication with retrieval of dental chair
"Rinse position".
Thecontrollerthatdeliverswatertothebowlwhenthedentalchair"Rinse
position"buttonispressedcanbeturnedon/o.
Thismodicationismadebyturningontheoperatingunitsimultaneously
holdingdownthedentalchairRinsingPositionandBowlWaterDelivery
buttons on the assistant’s board.
A beep sounds to signal the selected operating mode:
• 1 BEEP: controller enabled.
• 3 BEEPS: controller disabled.
NOTE: the controller is enabled by default.
Water to bowl controller modication with retrieval of dental chair
"Automatic return".
Thecontrollerthatdeliverswatertothebowlwhenthedentalchair"Auto-
maticreturn"buttonispressedcanbeturnedon/o.
Thismodicationismadebyturningontheoperatingunitsimultaneously
holdingdownthedentalchairAutomaticReturnandBowlWaterDelivery
buttons on the assistant’s board.
A beep sounds to signal the selected operating mode:
• 1 BEEP: controller enabled.
• 2 BEEPS: controller disabled.
NOTE: the controller is enabled by default.
Taking o the bowl lter and rinse spout.
• Pull up the spout ( l )andtakeito.
•Pullupthelter(q ) and its cover ( p ) to remove them.
•Turnthebowl(m ) counter-clockwise to release it and then pull it up to
take it out.
Disinfecting and cleaning.
WARNING!
Always wear gloves to prevent contact with infected material when
cleaning the bowl and bowl lter.
Thepartsaretobecleaneddailyattheendofeachworkday.
• Ceramic spouts and bowl: thoroughly wash with a specially formulated
cleaner(forexampleMD550OrotolDÜRR).
•Bowllter:cleanwithrunningwaterandcommercially-availablecleaning
products.
WARNING!
Do not use acids or harsh products.